Other, other text: additional information is provided in g.1., h.2, h.3., and h.6.Device evaluation: one device was received.A visual inspection found a scratched front cover, discolored line cord, cracked water tank, corroded quick connect and the stand off was missing behind the lcd.During the functional testing, the customer reported problem was unable to be duplicated but other problems were found.The lcd was missing a standoff and the water tank leaked.The pcb was replaced as well as the front cover and global display label, line cord, water tank cover, and quick connect.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.For all enquiries or follow-up questions related to the record, do not use regulatory.Responses@smiths-medical.Com located in sections g.1., please direct those to the following: regulatory.Responses@icumed.Com.
|