Additional information was added to d9, h3, h6 and h10.H10: the device was received for evaluation.A visual inspection with the naked eye and magnification noted damage to the patient adapter/tubing near the heat seal.Functional testing including leak, clear passage, and clamp function testing were performed with no issues noted.The reported condition leak was not verified, however, the condition of damage was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|