Model Number PRT-01201-000 |
Device Problems
Connection Problem (2900); Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/31/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the device needed service.During the device evaluation, ventec observed that the ventilator was unable to maintain peep (positive end expiratory pressure) and that its exhaled tidal volume (vte) levels were low.Additionally, the device was displaying a patient circuit disconnect alarm.There was no patient involvement associated with the reported event.
|
|
Manufacturer Narrative
|
H6: the ventilator was evaluated by ventec where the reported issues of it being unable to maintain peep (positive end expiratory pressure), having low exhaled tidal volume (vte) levels and displaying a patient circuit disconnect alarm were confirmed.Ventec replaced the internal flow transducer (ift) to resolve the reported issues.Proper device operation was then observed through functional and performance testing.The investigation determined that the cause of the reported issues was the ift.
|
|