• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH INNER SHEATH, FOR 26 FR. OUTER SHEATH; RESECTOSCOPE SHEATH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH INNER SHEATH, FOR 26 FR. OUTER SHEATH; RESECTOSCOPE SHEATH Back to Search Results
Model Number A22040A
Device Problem Scratched Material (3020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported to olympus that the inner sheath, for 26 fr.Outer sheath had the sheath connection loose.There were no reports of patient harm.The device was returned for evaluation and during the evaluation the following reportable malfunction was found: the advanced brake was scratched inside.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Manufacturer Narrative
The device was returned to olympus for evaluation, and the customer¿s allegation was confirmed.The evaluation found the following reportable issues: the advanced brake was scratched inside.The following non-reportable malfunctions were found during the device evaluation: the seal rubber was discolored.The investigation is ongoing and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 3 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of the reported event could not be identified.However, it¿s likely the cause is related to thermally induced impact, wear and tear, improper handling, mechanical overload like fall, shock, or similar stress.The suggested event is detectable and preventable by handling the device in accordance with the following section of the instructions for use: 4 before use: warning infection control risk properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.4.1 inspection and testing inspecting the product visually inspect the product.Make sure that it has: -- no corrosion -- no dents -- no scratches ceramic insulation at distal end visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).Warning risk of injury impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product if the product is damaged or does not function properly, contact an olympus representative or an authorized service center.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNER SHEATH, FOR 26 FR. OUTER SHEATH
Type of Device
RESECTOSCOPE SHEATH
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18177840
MDR Text Key328971757
Report Number9610773-2023-03354
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761029339
UDI-Public04042761029339
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K931994
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22040A
Device Lot Number21305
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2023
Was the Report Sent to FDA? No
Date Manufacturer Received04/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/12/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-