Model Number 8668
Device Problems Poor Quality Image (1408); Device Displays Incorrect Message (2591); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/04/2023
Event Type  malfunction  
Event Description
Reportable based on device analysis completed on (b)(6) 2023.It was reported that visualization issues occurred.An opticross hd imaging catheter was advanced for ultrasound examination of the target lesion, but the device could not show the image and first scope was not recognized.The procedure was completed with a different device.There were no patient complications reported and the patient condition following the procedure was stable.However, device analysis revealed tip detachment.
 
Manufacturer Narrative
The device was returned for analysis.During visual inspection sheath kinked and tip detached tip.Impedance testing showed an electrical open distal wave form between 0-10 cm from the distal housing the transducer level impedance test with the microprobe could not be performed due to the condition of the transducer/distal housing.Microscope inspection revealed tip detached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTICROSS HD
Type of Device
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18178837
MDR Text Key328634717
Report Number2124215-2023-64398
Device Sequence Number1
Product Code OBJ
Combination Product (y/n)N
Reporter Country CodeHR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 11/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/09/2024
Device Model Number8668
Device Catalogue Number8668
Device Lot Number0031202933
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/10/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-