Model Number LZ226822320S20 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/25/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
The device has not yet been made available for evaluation.Should further information or the device become available, a follow-up report will then be issued.
|
|
Event Description
|
Provider alleges the consumer allegedly smelled smoke while sitting in the radiance, but couldn't identify where it was coming from - the consumer allegedly found the chair on fire shortly after this, dragged it outside where it allegedly became engulfed in flames.
|
|
Event Description
|
Provider alleges the consumer allegedly smelled smoke while sitting in the radiance, but couldn't identify where it was coming from - the consumer allegedly found the chair on fire shortly after this, dragged it outside where it allegedly became engulfed in flames.
|
|
Manufacturer Narrative
|
Remnant harnessing was examined and no evidence of arcing was found.With this, the origin of the event is unknown.
|
|