Catalog Number 05942861001
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/11/2023
Event Type  malfunction  
Manufacturer Narrative
The test strips were requested for investigation.The product has not been received at this time.If the product is returned in the future, a follow up report will be submitted.On a regular basis accu-chek inform strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.
 
Event Description
There was an allegation of questionable results from accuchek inform ii serial number (b)(6) and compared to a laboratory result.At 8:40 p.M., the meter result was 43 mg/dl.At 9:06 p.M., the meter result was 334 mg/dl.At 9:08 p.M., the meter result was 80 mg/dl.At an unknown time, the meter result was 68 mg/dl.At an unknown time, the laboratory result was 62 mg/dl.The patient received dextrose based on the results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCU-CHEK INFORM II TEST STRIPS
Type of Device
BLOOD GLUCOSE MONITORING TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18182707
MDR Text Key329617346
Report Number1823260-2023-03728
Device Sequence Number1
Product Code LFR
UDI-Device Identifier00365702428102
UDI-Public00365702428102
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K121679
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05942861001
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/03/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
DEXTROSE 10% IV
-
-