Model Number ILS-1000-CS |
Device Problems
Device Displays Incorrect Message (2591); Application Program Problem (2880); Data Problem (3196)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 11/03/2023 |
Event Type
Injury
|
Event Description
|
According to the reporter, during a procedure, the user received multiple "load error" messages that prevented the surgeon from proceeding beyond the sensor triplet confirmation screens.Following attempts to resolve were made: backing out and re-planning, continuing to the procedure from planning, starting a new procedure and selecting the patient from the 'display patient list', re-starting the system completely, and re-burning the cd.None of the attempts were able to resolve the issue and the procedure was aborted.
|
|
Manufacturer Narrative
|
H6 patient codes - e2402 (cancelled case) out of an abundance of caution, superdimension is filing this mdr due to the additional risk associated with multiple exposures to general anesthesia.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a procedure, the user received multiple 'load error' messages that prevented the surgeon from proceeding beyond the sensor triplet confirmation screens.The computed tomography scan and plan were unable to load.Following attempts to resolve were made: backing out and re-planning, continuing to the procedure from planning, starting a new procedure and selecting the patient from the 'display patient list', re-starting the system completely, and re-burning the cd.None of the attempts were able to resolve the issue, and the procedure was aborted.
|
|
Manufacturer Narrative
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted that the ct reconstruction parameters were outside the recommendation (a load error).It was re ported that the user received multiple "load error" messages that prevented the surgeon from proceeding beyond the sensor triplet confirmation screens and the computed tomography scan and plan were unable to load.The reported issues were confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|