Continuation of d10: product id 97755 lot# serial# (b)(6) implanted: explanted: product type recharger section d information references the main component of the system.Other relevant device(s) are: product id: 97755, serial/lot #: (b)(6), ubd: , udi#: (b)(4) h3: analysis of the 97755 intellis recharger (s/n (b)(6)) revealed that "no device found" message was shown.This regulatory report is being submitted as part of a retrospective review as part of remediation plan 411.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
Information was received from a patient (pt) regarding an external device.The reason for call was since about 3 days ago, pt mentioned their recharger antenna cord was getting hot to the touch when they charged their implanted neurostimulator( ins); pt said they had to hold their recharger antenna coil in "exactly the right position" to get their recharger antenna to connect to their ins to charge.Pt said the entered passive recharge mode at one point in time and the middle number changed from 0 to 90 "percent" when they moved the recharger antenna very slightly.Pt said "it was like a isometric exercise" to get the ins to charge.During the call, the pt turned the relay box over to read the serial number and the recharger antenna disconnected from the ins while charging the ins.Recharger antenna was very sensitive.Ins charge was "low." pt said their program they used was working well, but they needed to charge their ins every other day.No symptoms or patient complications were reported.
|