• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ISLEEVE 14F; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ISLEEVE 14F; INTRODUCER, CATHETER Back to Search Results
Model Number 10445
Device Problems Difficult to Remove (1528); Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/25/2023
Event Type  malfunction  
Event Description
It was reported that failure to advance within the artery occurred.The patient length was 130cm.Vascular access was obtained via a transfemoral approach.The vessel diameter was 6.2 min/7.0 max with moderate tortuosity and mild calcifications.A 14f isleeve introducer sheath was inserted and was unable to be fully advanced into the patient.An unknown manufacturer's valve delivery system was advanced through the 14f isleeve introducer sheath.The unknown manufacturer's valve was implanted to treat the native aortic annulus.The unknown manufacturer's valve delivery system was removed with the 14f isleeve introducer sheath.A new 14f isleeve introducer sheath was advanced to complete the procedure.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISLEEVE 14F
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18187318
MDR Text Key328998031
Report Number2124215-2023-65787
Device Sequence Number1
Product Code DYB
UDI-Device Identifier08714729950660
UDI-Public08714729950660
Combination Product (y/n)N
Reporter Country CodeCI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10445
Device Catalogue Number10445
Device Lot Number0031693880
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/25/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/25/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-