H4: device manufactured on october 31, 2022 - november 1, 2022.H10: one actual sample and a photograph was received for evaluation.A visual inspection with the unaided eye was performed on the actual sample with no issues noted.Functional testing was performed by installing the sample onto a compounder.The sample was evaluated by reproducing the customer complaint as closely as possible by attaching a lipid solution to ports 1 through 23 and sterile water to port 24 for testing.The valve set was then analyzed at various points throughout the prime, verify process, and pump calibration process and no leak was identified on any of the inlet ports.The photograph that was provided appeared to have lipid entrained into the tubing from the valve set end.Therefore, the reported condition was not verified during the actual sample evaluation, however, based on the sample photo the leak was verified.The cause of the condition could not be determined.A nonconformance has been opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|