It was reported to boston scientific corporation that a resolution 360 clip device was used during a colonoscopy procedure performed on october 26, 2023.During the procedure, when the clip was passed down the working channel, the physician attempted to clip the lesion however, the clip did not deploy from the catheter.The technician noticed that the clip was difficult to close so the physician attempted to deploy the clip but was unsuccessful.The clip was eventually pulled off of the tissue while still attached to catheter.There was abnormally high resistance felt before the handle spool reached the end of the stroke.The procedure was completed with a non-bsc device.The photos submitted by the customer shows that the control wire was broken.There were no patient complications reported as a result of this event.
|