Model Number M00521232
Device Problems Break (1069); Activation, Positioning or Separation Problem (2906); Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/26/2023
Event Type  malfunction  
Manufacturer Narrative
Block h6: imdrf device code a15 captures the reportable event of clip did not deploy.
 
Event Description
It was reported to boston scientific corporation that a resolution 360 clip device was used during a colonoscopy procedure performed on october 26, 2023.During the procedure, when the clip was passed down the working channel, the physician attempted to clip the lesion however, the clip did not deploy from the catheter.The technician noticed that the clip was difficult to close so the physician attempted to deploy the clip but was unsuccessful.The clip was eventually pulled off of the tissue while still attached to catheter.There was abnormally high resistance felt before the handle spool reached the end of the stroke.The procedure was completed with a non-bsc device.The photos submitted by the customer shows that the control wire was broken.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTION 360 CLIP
Type of Device
HEMOSTATIC METAL CLIP FOR THE GI TRACT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
propark, coyol
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18191793
MDR Text Key328801878
Report Number3005099803-2023-06247
Device Sequence Number1
Product Code PKL
UDI-Device Identifier08714729875642
UDI-Public08714729875642
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K222503
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 11/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00521232
Device Catalogue Number54774
Device Lot Number0032311956
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/26/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/29/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-