DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Moisture or Humidity Problem (2986)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/14/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that, the cs300 intra-aortic balloon pump (iabp) unit is leaking moisture coming from the fuse.There was no patient harm.
|
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that prior to use, the cs300 intra-aortic balloon pump (iabp) unit is leaking moisture coming from the fuse.There was no patient harm.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to investigate the issue.Upon inspection, the fse performed condensation removal procedure due to condensation in the line.The fse confirmed that the iabp failed the safety disk leak test.So, in order to fix the issue, the fse replaced the safety disk.Additionally, the fse then verified that the unit calibrated and passed all functional and safety checks to factory specifications.The unit was then returned to the customer and cleared for clinical use.
|
|
Event Description
|
N/a.
|
|