ICU MEDICAL COSTA RICA LTD. PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 103 INCH; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 146870489 |
Device Problem
Particulates (1451)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/01/2023 |
Event Type
malfunction
|
Event Description
|
The incident involved a primary plum set, clave secondary port, clave y-site, secure lock, 103 inch on an unknown date.A nurse had identified contaminated iv tubing prior to using it with a patient.The product was isolated and kept in the product packaging.The product was sealed.The particulate was noted once opened, and the nurse was going to use it.The product was stored in a clean, temperature-controlled storage room on shelf in a hard plastic bin with like products.There was no patient involvement or harm.
|
|
Manufacturer Narrative
|
The device has been requested to be returned for evaluation; however, it has not yet been received.
|
|
Manufacturer Narrative
|
Received one open/unused 146870489 primary plum set for inspection.Embedded burnt material was observed in the wall of the sight chamber.The sight chamber was cut open and the particles were confirmed to be embedded in the wall, not in the fluid path.The most probable cause is burnt material embedded in the wall of the sight chamber during the molding process/manufacturing.While the device fails visual inspection, the embedded material is not in the fluid path, and does not affect the functionality of the device.The lot history was reviewed and no nonconformities were identified that may have contributed to the reported complaint.
|
|