Catalog Number 146870489
Device Problem Particulates (1451)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2023
Event Type  malfunction  
Event Description
The incident involved a primary plum set, clave secondary port, clave y-site, secure lock, 103 inch on an unknown date.A nurse had identified contaminated iv tubing prior to using it with a patient.The product was isolated and kept in the product packaging.The product was sealed.The particulate was noted once opened, and the nurse was going to use it.The product was stored in a clean, temperature-controlled storage room on shelf in a hard plastic bin with like products.There was no patient involvement or harm.
 
Manufacturer Narrative
The device has been requested to be returned for evaluation; however, it has not yet been received.
 
Manufacturer Narrative
Received one open/unused 146870489 primary plum set for inspection.Embedded burnt material was observed in the wall of the sight chamber.The sight chamber was cut open and the particles were confirmed to be embedded in the wall, not in the fluid path.The most probable cause is burnt material embedded in the wall of the sight chamber during the molding process/manufacturing.While the device fails visual inspection, the embedded material is not in the fluid path, and does not affect the functionality of the device.The lot history was reviewed and no nonconformities were identified that may have contributed to the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 103 INCH
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ICU MEDICAL COSTA RICA LTD.
zona franca global
la aurora heredia
CS 
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key18193684
MDR Text Key329093059
Report Number9615050-2023-00669
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10887787005650
UDI-Public(01)10887787005650(17)260501(10)13579625
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative,Distributor
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number146870489
Device Lot Number13579625
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-