Catalog Number 383536
Device Problem Delivered as Unsterile Product (1421)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2023
Event Type  malfunction  
Manufacturer Narrative
The complaint of a defective device was confirmed and the cause was packaging related.One 22g nexiva unit from lot #3116955 was provided for investigation.The nexiva catheter was caught between the sealing surfaces of the top and bottom web packaging, which created gaps in the seal.The appropriate manufacturing personnel were notified of this complaint.A device history record review showed no non-conformances associated with this issue during the production of this batch.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Reporter facility full name: (b)(6) medical center elk.
 
Event Description
It was reported that bd nexiva the following information was provided by the initial reporter: iv came out of packaging with manufacturing defects.See picture from customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD NEXIVA
Type of Device
PERIPHERAL IV CATHETERS
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18196669
MDR Text Key329895197
Report Number1710034-2023-01337
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier30382903835363
UDI-Public(01)30382903835363
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102520
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number383536
Device Lot Number3116955
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/26/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-