The leaks or loose pipes at the y connector would interrupt the patient therapy, and the patient could become hypoxia no images or returned inventory provided, complaint is unconfirmed.The issue is stated as the cannula is leaking.There was nothing in inventory to review as well as nothing out of the ordinary for the dhr's.In the last 24 months, there have been 2 related complaints.Root cause cannot be determined at this time.However, likely root cause is little or no bonding agent applied to bonding location during assembly.Risk assessment: for patient risk, rma-20017f, step r87 issue is stated as "low oxygen delivery", potential hazard being "loss/delay in oxygen therapy leading to hypoxia", causation "cannula tubing deteriorates resulting in leaks prior to first use", s=8, o=1, rpn=8 (less than 25, therefore risk is acceptable).For regulatory risk the lack of directly related complaints indicates a low-risk.
|