On 9th november, 2023 getinge became aware of an issue with one of surgical equipment - configuration of module mobili 800 modutec and 1 drawer 400x500 modutec.It was found during annual maintenance the paint has worn off in the corner of the drawer tableau.The issue was confirmed by photographic evidence.It was established by technician the issue occurred probably due to bumping and cleaning.There was no injury reported, however, we decided to report the issue in abundance of caution as any particles falling off into sterile field or during procedure may cause contamination.
|