SHIRAKAWA OLYMPUS CO., LTD. ULTRASONIC PROBE Back to Search Results
Model Number UM-G20-29R
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2023
Event Type  malfunction  
Event Description
The customer reported it was difficult to see the ultrasound images using the ultrasonic probe.They adjusted the settings and used it to complete the unspecified procedure.The device was returned for evaluation.During the device evaluation, leakage of the ultrasound media was found.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the malfunction found during evaluation.
 
Manufacturer Narrative
E1: establishment name: (b)(6).The device was returned to olympus for evaluation and the customer's allegation was confirmed.In addition to the findings reported in b5, the device evaluation found there was a hole, scratches and indentations on the tip sheath.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.The presence of holes in the tip sheath indicates that some stress was applied to the tip sheath.As the internal ultrasound medium leaked out from the hole and decreased, ultrasound could not be transmitted normally and ultrasound images could not be visualized properly, leading to the phenomenon that the ultrasound images were difficult to see.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC PROBE
Type of Device
ULTRASONIC PROBE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18215875
MDR Text Key330456476
Report Number3002808148-2023-13349
Device Sequence Number1
Product Code ITX
UDI-Device Identifier04953170368578
UDI-Public04953170368578
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K944610
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUM-G20-29R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2023
Was the Report Sent to FDA? No
Date Manufacturer Received12/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/12/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-