Model Number V60 V60PLUS VENTILATOR
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/31/2023
Event Type  malfunction  
Event Description
Philips received a complaint by the biomedical engineer (bme) on the v60 indicating that the fan was pulsating with power when charging.It was reported that there was no patient involvement at the time the issue was discovered.The biomedical engineer (bme) reported to the remote service engineer (rse) that the fan was pulsating with power when charging.The rse troubleshot the device with the bme and found that the battery voltage was 14 volts after the device was plugged in for a while.The rse advised the bme to charge the battery for the rest of the day-- if the voltage was still low, the rse recommended that the bme should replace the battery and provided the part number of the replacement battery to the customer for repair upon request.The investigation is ongoing.
 
Manufacturer Narrative
The biomedical engineer (bme) reported to the remote service engineer (rse) that the fan was pulsating with power when charging.The rse troubleshot the device with the bme and found that the battery voltage was 14 volts after the device was plugged in for a while.The rse advised the bme to charge the battery for the rest of the day if the voltage was still low, the rse recommended that the bme should replace the battery and provided the part number of the replacement battery to the bme for repair upon request.Multiple attempts have been made to try to obtain further information about this case, but no response was received from the bme.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18217747
MDR Text Key329135733
Report Number2518422-2023-32023
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/01/2023
Date Device Manufactured12/19/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-