Model Number P3700A004164
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/13/2023
Event Type  malfunction  
Manufacturer Narrative
The baxter technician found the brakes needed to be adjusted.The affinity® 4 birthing bed requires an effective maintenance program.We recommend that you do annual preventive maintenance (pm) for joint commission certification.Pm not only meets joint commission requirements but can help make sure of a long, operative life for the affinity® 4 birthing bed.Two effective ways to reduce downtime and make sure the patient remains comfortable are to keep accurate records and maintain the affinity® 4 birthing bed.Check the tires for cuts, wear, tread life, etc.Apply the brake, and check to ensure that the bed will not move.If the bed moves, inspect it for wear, and adjust if required.Apply the steering pedal and check the steering to ensure proper locking action when activated.A search of the baxter maintenance records showed baxter performed preventative maintenance on this bed on july 2023.It is unknown if the facility performed any other preventative maintenance on this bed.The technician adjusted the brakes to resolve the reported event.Based on this information, no further action is required.
 
Event Description
Baxter received a report from a baxter technician stating the brakes had an issue and were not holding.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITY THREE BED FRAME
Type of Device
TABLE, OBSTETRIC (AND ACCESSORIES)
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
brad wheeler
1069 state route 46 east
batesville, IN 47006
3128199307
MDR Report Key18219642
MDR Text Key329151249
Report Number1824206-2023-01323
Device Sequence Number1
Product Code KNC
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K915779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP3700A004164
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/05/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-