Patient (b)(6) was at (b)(6) hospital in (b)(6) for a battery change of her item number 3116 which was implanted on (b)(6) 2014.I checked her device before implant and noticed that the impedance of the system was over 800ohms.We were hoping by connecting the leads of her system into a new device, that the impedance issue would normalize and that her leads would not need to be explanted.After opening connecting her leads to a new 37800 enterra ii stimulator, the impedance reading was over 20000.Next step decided by dr.(b)(6) was to remove the existing leads and replace them with a set of new leads.During the process of removing the electrodes from the surface of the stomach, the stomach perforated with an opening to the inside.As a result of a hole being created in the wall of the stomach, dr.(b)(6) decided to repair the hole and allow the patient to heal for a few weeks before replacing the enterra system.
|