Model Number WA53005A
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/09/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was not confirmed.The device evaluation findings are as follows: the image was shadowy and there was foreign material inside the optical system, the identification on the eyepiece was worn, and the eyepiece funnel was aging.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus, the olympus rigid scope had crystals on the eyepiece.The issue was found during reprocessing.The procedure was completed with a similar device.There was no delay, the patient was not under anesthesia, and there was no report of patient injury or medical intervention associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 10 MM, 30°, HD, QUICK LOCK, AUTOCLAVABLE
Type of Device
RIGID SCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18222382
MDR Text Key329434333
Report Number9610773-2023-03404
Device Sequence Number1
Product Code EWY
UDI-Device Identifier04042761052764
UDI-Public04042761052764
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K950076
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA53005A
Device Lot Number735416
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/12/2023
Was the Report Sent to FDA? No
Date Manufacturer Received11/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/22/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
VISERA ELITE VIDEO SYSTEM CENTER
-
-