Model Number N/A |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/17/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the switch is broken.The event occurred outside of surgery.There was no reported patient harm, or delay.Due diligence is complete, no further information is available at this time.
|
|
Manufacturer Narrative
|
This event is recorded with zimmer biomet under (b)(4).This medwatch is being filed as an initial report e1 phone: (b)(6) g2 foreign: south korea customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the switch was broken.The switch and switch mount were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported that the switch is broken.The event occurred outside of surgery during kit inspection.There was no reported patient harm, or delay.Due diligence is complete, no further information is available at this time.
|
|