Model Number AZURION 7 B20
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/20/2023
Event Type  malfunction  
Event Description
It has been reported to philips that the c-arm cover fell off.The issue was found during clinical use.No harm has been reported to philips.Philips has started an investigation of this complaint.
 
Manufacturer Narrative
Philips has investigated this complaint.The philips field service engineer (fse) inspected the system onsite and confirmed that the c-arm cover had fallen off.During troubleshooting, the fse attempted to re-attach the cover but found that it couldn't be reattached.The fse handed over the system to customer with limitations.The fse fitted a new cover for the lateral tube in the follow up work order.After replacing the cover, the system was returned to use in good working order.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZURION
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
derrick massuri
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key18226377
MDR Text Key329215217
Report Number3003768277-2023-06591
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00884838085367
UDI-Public00884838085367
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K172822
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/29/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAZURION 7 B20
Device Catalogue Number722068
Was Device Available for Evaluation? Yes
Date Manufacturer Received11/20/2023
Date Device Manufactured12/20/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-