SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA BREATHING CIRCUIT BAG; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number 670001 |
Device Problem
Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Other text: d4: lot number, expiration date and h4: manufacture date are unknown, no information has been provided to date.G5: 510k is bank, device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that during pre-testing, the bag was torn by the sharp part inside the black connector ring.No adverse effects reported.
|
|
Manufacturer Narrative
|
Updated two photos and one device were returned for evaluation.The two photos show a tear in the breathing bag.The returned sample was visually inspected and detected a tear also.The reported complaint is confirmed.The root cause cannot be determined and will continue to be monitored and further actions taken accordingly.A review of the device history records could not be completed as no lot number was provided by the customer.
|
|