• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE, INC. CLOSURE TOP; VITALITY SPINAL FIXATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE, INC. CLOSURE TOP; VITALITY SPINAL FIXATION SYSTEM Back to Search Results
Catalog Number 07.02010.001
Device Problem Migration (4003)
Patient Problems Pain (1994); Deformity/ Disfigurement (2360); Insufficient Information (4580)
Event Type  Injury  
Event Description
It was reported that two closure tops migrated out of their mating iliac screws post-operatively.A revision surgery is scheduled.This is report two of four for this event.
 
Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Manufacturer Narrative
If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.Reference reports 3012447612-2023-00375 through 3012447612-2023-00378.
 
Event Description
It was reported that two closure tops migrated out of their mating iliac screws post-operatively.A revision surgery was performed in which the closure tops were removed and replaced while the iliac screws remained implanted.The patient's pain and posture improved following the revision surgery.The initial construct was t-11 to ilium.This is report two of four for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLOSURE TOP
Type of Device
VITALITY SPINAL FIXATION SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer Contact
sabrina abla
10225 westmoor dr.
westminster, CO 80021
7206965158
MDR Report Key18232860
MDR Text Key329293581
Report Number3012447612-2023-00376
Device Sequence Number1
Product Code NKB
UDI-Device Identifier00889024003170
UDI-Public(01)00889024003170(10)W6873411(11)210407
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150896
Exemption Number5645646
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/29/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number07.02010.001
Device Lot NumberW767911
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/06/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexMale
Patient EthnicityNon Hispanic
-
-