SHIRAKAWA OLYMPUS CO., LTD. DUODENOFIBERSCOPE FOR ULTRASONIC SURVEY; ULTRASOUND GASTROVIDEOSCOPE
|
Back to Search Results |
|
Model Number TGF-UC260J |
Device Problem
Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned and evaluated, and the customer¿s allegation was confirmed.In addition to the malfunction documented in b5, the additional evaluation findings are as follows: due to deformation of u/d (up/down) knob, water tightness was lost, due to a pinhole on ch (channel) tube, water tightness was lost, due to wear of angle wire, bending angle in up direction did not meet the standard value, due to wear of angle wire, the play of u/d knob was out of the standard value, adhesive on a-rubber was detached, and the ultrasonic transducer had corrosion.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Event Description
|
The customer reported to olympus, that the duodenofiberscope had an air leak.The device was returned for evaluation.During the device evaluation, it was found that the probe was chipped.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation (h6/10).A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause could not be conclusively determined.The following information from the instructions for use (ifu) pertains to this event: "warning do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, or endoscope connector.Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, or endoscope connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient." olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|