Model Number 97715
Device Problems Failure to Interrogate (1332); Use of Device Problem (1670); Delayed Charge Time (2586); Malposition of Device (2616); Charging Problem (2892); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/01/2023
Event Type  malfunction  
Manufacturer Narrative
This regulatory report is being submitted as part of a retrospective review as part of remediation plan 411.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a patient (pt) who was implanted with an implantable neurostimulator (ins) for unknown indications for use.The reason for call was that they couldn't get the ins battery charged.Pt said that they couldn't get the recharger to register with the controller.Pt said that when the ins was charging the controller light blinked green and recharging excellent would be indicated.Pt said that they had been recharging the ins since seven this morning and the ins started out at 20% and the ins was only at 30% now.Pt said that if they moved their hand off of the recharger then the ins would quit charging.Pt said that they didn't have this problem with their last ins so they thought their current ins was implanted too deep.Pss had the pt remove the battery pack from the controller and connect the controller to the ac power supply; pt confirmed that the controller powered on.Pss then had the pt reinsert the battery pack into the controller while the controller was still connected to the ac power supply; pt confirmed seeing the controller light flashing green.Pss had the pt unlock the controller; the controller was at 80% with recharging indicated and the ins was at 30%.Pt saw no apparent damage to the recharger.Pss had the initiate an ins recharging session.Pt said that the recharger wouldn't stay on the implant so their wife was holding the recharger over the ins for them.Pt said that they could never find the ins.Pt saw recharging excellent with the controller light flashing green.Pt then saw poor recharge quality; pss reviewed screen meaning with the pt.Pt said that they finally found the ins when they first started using the device and so they had taken a pencil and drew around the paddle when recharging excellent was indicated but since the markings had faded out.Pt said again that recharging excellent would be indicated but if they moved their hand then the ins quit charging.Pss understood that the recharger was not staying in place properly over the ins.Pt saw recharging excellent appear and the equipment stayed on recharging excellent for the rest of the call.Pss redirected to hcp to further address the issue as the equipment was working as intended during the call.Pt said that they had been having this trouble ever since their current ins was implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key18234218
MDR Text Key330564141
Report Number3004209178-2023-24496
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00763000315467
UDI-Public00763000315467
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/29/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/14/2023
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? No
Date Manufacturer Received05/16/2023
Date Device Manufactured09/16/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
"SEE H10...."
Patient Age81 YR
Patient SexMale
-
-