OLYMPUS WINTER & IBE GMBH TELESCOPE, 10 MM, 0°, HD, QUICK LOCK, AUTOCLAVABLE; THORACOSCOPE SYSTEM
|
Back to Search Results |
|
Model Number WA53000A |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/09/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned and evaluated, and the customer¿s allegation was confirmed.In addition to the eyepiece lens being damaged, additional findings were as follows: the outer tube was skewed with dents; light guide fibers were damaged; particles were inside the device; the serial number was worn; and there was no color ring.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
|
|
Event Description
|
The customer reported to olympus that the telescope, 10 mm, 0°, hd, quick lock, autoclavable had a blurry image.The event occurred during reprocessing.There was no patient harm associated with the event.The device was evaluated, and it was found that the eyepiece lens was damaged.This mdr (medical device report) is being submitted to capture the reportable malfunction found during the device evaluation.
|
|