OLYMPUS WINTER & IBE GMBH TELESCOPE, 10 MM, 0°, HD, QUICK LOCK, AUTOCLAVABLE; THORACOSCOPE SYSTEM Back to Search Results
Model Number WA53000A
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/09/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned and evaluated, and the customer¿s allegation was confirmed.In addition to the eyepiece lens being damaged, additional findings were as follows: the outer tube was skewed with dents; light guide fibers were damaged; particles were inside the device; the serial number was worn; and there was no color ring.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
 
Event Description
The customer reported to olympus that the telescope, 10 mm, 0°, hd, quick lock, autoclavable had a blurry image.The event occurred during reprocessing.There was no patient harm associated with the event.The device was evaluated, and it was found that the eyepiece lens was damaged.This mdr (medical device report) is being submitted to capture the reportable malfunction found during the device evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 10 MM, 0°, HD, QUICK LOCK, AUTOCLAVABLE
Type of Device
THORACOSCOPE SYSTEM
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18236354
MDR Text Key329431664
Report Number9610773-2023-03435
Device Sequence Number1
Product Code EWY
UDI-Device Identifier04042761052757
UDI-Public04042761052757
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K912362
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA53000A
Device Lot Number637951
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2023
Was the Report Sent to FDA? No
Date Manufacturer Received11/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/07/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CV-S190
-
-