H10: h3, h6: the reported device was received for evaluation.Visual inspection observed the two returned instruments show no manufacturing abnormalities.The electrodes have not been used.Both devices were out of the original packaging.No packaging returned.A functional evaluation showed both the opened devices were plugged into the controller and registered settings (0,2).When used with a bypass box, both the wands were able to generate plasma and coagulation as intended.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the activity test workmanship found that the operator must verify the default setting and should reject units that do not default to the correct settings.The root cause can be attributed to the user.Factors that could have contributed to the reported event include previous use and connecting and disconnecting the wand.No containment or corrective actions are recommended at this time.
|