Model Number LIFEVEST WCD 4000 SYSTEM
Device Problem Battery Problem (2885)
Patient Problem Lethargy (2560)
Event Date 11/14/2023
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of battery pack has been completed.The reported problem (will not power up) was confirmed due to a loose bus bar inside of the battery.The root cause of the loose busbar cannot be positively identified.There was no adverse event that resulted from the damaged battery.
 
Event Description
A us distributor reported that a battery was unable to power on a monitor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA
Manufacturer Contact
zoll mfg corporation
121 gamma drive
pittsburgh, PA 15238
MDR Report Key18238244
MDR Text Key329380156
Report Number3008642652-2023-12159
Device Sequence Number1
Product Code MVK
Combination Product (y/n)N
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 11/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberLIFEVEST WCD 4000 SYSTEM
Date Manufacturer Received11/29/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-