• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS GLUCOSE HK GEN.3; GLUCOSE HEXOKINASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS GLUCOSE HK GEN.3; GLUCOSE HEXOKINASE Back to Search Results
Catalog Number 08057800190
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/10/2023
Event Type  malfunction  
Manufacturer Narrative
The serial number of the c 503 analyzer is (b)(6).The investigation is ongoing.
 
Event Description
The initial reporter stated they received discrepant results for one patient sample tested with glucose hk gen.3 on a cobas pro c 503 analytical unit.No questionable results were reported outside of the laboratory.The sample initially resulted in a glucose value of 12.7 mg/dl.The value did not agree with the patient's clinical status, so it was repeated.The sample was repeated on a second analyzer, resulting in a glucose value of 85.6 mg/dl.The repeat value was deemed correct as it matched the patient's clinical status.
 
Manufacturer Narrative
The field service engineer found no issues with the analyzers.Precision studies were performed and were acceptable.Controls and a hardware check passed.The investigation could not identify a product problem.The cause of the event could not be determined.The issue was resolved after the service actions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLUCOSE HK GEN.3
Type of Device
GLUCOSE HEXOKINASE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18239958
MDR Text Key329396423
Report Number1823260-2023-03808
Device Sequence Number1
Product Code CFR
UDI-Device Identifier07613336121283
UDI-Public07613336121283
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K191899NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number08057800190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/22/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age56 YR
Patient SexMale
-
-