Catalog Number 08057800190 |
Device Problems
Incorrect, Inadequate or Imprecise Result or Readings (1535); Low Test Results (2458)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/10/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The serial number of the c 503 analyzer is (b)(6).The investigation is ongoing.
|
|
Event Description
|
The initial reporter stated they received discrepant results for one patient sample tested with glucose hk gen.3 on a cobas pro c 503 analytical unit.No questionable results were reported outside of the laboratory.The sample initially resulted in a glucose value of 12.7 mg/dl.The value did not agree with the patient's clinical status, so it was repeated.The sample was repeated on a second analyzer, resulting in a glucose value of 85.6 mg/dl.The repeat value was deemed correct as it matched the patient's clinical status.
|
|
Manufacturer Narrative
|
The field service engineer found no issues with the analyzers.Precision studies were performed and were acceptable.Controls and a hardware check passed.The investigation could not identify a product problem.The cause of the event could not be determined.The issue was resolved after the service actions.
|
|
Search Alerts/Recalls
|