Catalog Number 8065753109
Device Problem Difficult to Open or Close (2921)
Patient Problem Insufficient Information (4580)
Event Date 05/24/2023
Event Type  malfunction  
Event Description
An ophthalmologist reported opening of an auto-fusion valve failed during prime.The surgery was completed after replacing an infusion tube with another one.The procedure type was vitrectomy.No information about patient harm.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The returned sample was visually inspected and no obvious defects were found.The non-invasive flow sensor (nifs) on the cassette housing was in good condition.A calibrated console representing the current software version was used to test the sample.The ball in the drip chamber¿s check valve moved freely per specification.The light-emitting diode (led) rings on the console turned green as the probe connectors were engaged to the console indicating the proper communication between the probe and the console.The sample could prime and pass intraocular pressure (iop) calibration successfully.The left and right measured crystal signal strength, pressure and flow rates were all met specification.No anomalies were observed during priming.No system message code was generated during testing.Fluid flowed from the balance salt solution (bss) bottle to the drain bag without any interfering.No leakage was detected from the pump elastomer or on the pump area of the fluidics module.Cleaning process was able to be performed after functional test had completed.The root cause of the customer's complaint could not be established; the returned sample functioned per specifications.No contributing factors could be identified that could cause the customer¿s experience.After an investigation of this complaint, it has been determined that this sample functioned per specifications; therefore, no corrective action is required at this time.No adverse trends have been observed associated with the reported product and event.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PACK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18240171
MDR Text Key329397897
Report Number1644019-2023-01735
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657531097
UDI-Public00380657531097
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065753109
Device Lot Number150N53
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/17/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-