HEARTSINE TECHNOLOGIES LTD PACKAGE, 350P,PP03J,JA,350-STR-JA-10; AUTOMATED EXTERNAL DEFIBRILLATOR
|
Back to Search Results |
|
Model Number 350P |
Device Problems
Failure to Power Up (1476); No Audible Prompt/Feedback (2282)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/06/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Heartsine has requested return of the device for investigation.Upon completion, the conclusions will be submitted in a follow-up report.
|
|
Event Description
|
The customer contacted heartsine to report that their device did not emit audio prompts upon start up.This issue has the potential to either delay, or prevent, defibrillation from being delivered.There was no report of patient use associated with the reported event.
|
|
Manufacturer Narrative
|
Heartsine's investigation of the device confirmed the reported fault as upon receipt the device was found to be unable to switch on.This fault was attributed to corrosion on the membrane tail due to being stored outside of indicated conditions.The device was scrapped by heartsine and the customer was provided with a replacement device.
|
|
Event Description
|
The customer contacted heartsine to report that their device did not emit audio prompts upon start up.This issue has the potential to either delay, or prevent, defibrillation from being delivered.There was no report of patient use associated with the reported event.
|
|