Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: the force bipolar instrument had poor energy output.The customer replaced the energy cable, but the issue was not resolved.It was determined that the force bipolar instrument was defective.The issue occurred during the procedure on (b)(6) 2023.There was no patient injury.The procedure was completed with a backup force bipolar instrument.There was no fragment falling into the patient¿s anatomy.Intuitive surgical, inc.(isi) received the force bipolar instrument involved with this complaint and completed the device evaluation.Failure analysis replicated and confirmed the customer reported complaint.The instrument was found to have a segment of the conductor wire dislodged from connector at the back end which likely causing energy delivery to not working properly.The instrument failed an electrical continuity test.After the dislodged conductor was plugged back into the connector to retest, the instrument passed electrical continuity.Visual inspection did not find any insulation damage, exposed wires, or thermal damage.Additional investigation found excessive amount of dried residue around the clamping pulley cable grooves.
|