Related manufacturing ref: 2184149-2023-00231.During a typical atrial flutter ablation procedure, a communication issue occurred and the procedure was cancelled.During the procedure, the amplifier encountered multiple disconnections, and higher than usual distortion that made it impossible to work in voxel mode.When switched to navx mode to move forward with case, distortion was higher than 3.Only after completely removing the x-ray, the system passed distortion.After that was done and after confirming that everything was connected, the system was unable to collect any magnetic data (no orange dot during the entire case).Check metal and baseline was performed, respiration was performed, prs dots were green, and the tactiflex dot was green.Se field scaling couldn¿t be completed as magnetic data could not be collected, and therefore no force was displayed on the catheter and the procedure was cancelled with no consequences to the patient.
|