CORDIS US CORP. TEMPO; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Catalog Number 451531H0
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/06/2023
Event Type  malfunction  
Manufacturer Narrative
The device is available for analysis but has not yet been received.Additional information is pending and will be submitted within 30 days upon receipt.
 
Event Description
As reported, the tip end of a 5f tempo sidewinder ii (sim2) diagnostic angiography catheter fractured during the procedure after entering the body and was removed from the patient.The patient was hospitalized for observation.The device is expected to be returned for evaluation.
 
Manufacturer Narrative
As reported, the tip end of a 5f tempo sidewinder ii (sim2) diagnostic angiography catheter fractured during the procedure after entering the body and was removed from the patient.The patient was hospitalized for observation.Without the return of the device or images for analysis, the reported customer event ¿brite tip/distal tip-cracked¿ could not be confirmed.Shipping, storage, or handling factors may have contributed to the reported event.Users are trained to inspect for signs of damage prior to and during use.Any product with damage is not to be used.Information for safety is provided in the products labeling with the intent to make the user aware of the risks.According to the instructions for use (ifu), although not intended as a mitigation of risk, ¿do not use if package is open or damaged.To prevent damage to the catheter tip during removal from the package, grasp the hub and withdraw the catheter.¿ based on the information available, there is no indication that the event is related to the device design or manufacturing process.Therefore, no preventive or corrective actions will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPO
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
CORDIS US CORP.
14201 nw 60th avenue
miami lakes, florida 33014
Manufacturer (Section G)
CORDIS US CORP.
14201 nw 60th avenue
miami lakes, florida 33014
Manufacturer Contact
karla castro
santiago troncoso 808
juarez, chihuahua 
7863138372
MDR Report Key18245705
MDR Text Key329470124
Report Number9616099-2023-06651
Device Sequence Number1
Product Code DQO
UDI-Device Identifier10705032008587
UDI-Public10705032008587
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973401
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number451531H0
Device Lot Number18222298
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/18/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/05/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
TEMPO 5F 100CM SIM2.
-
-