AIZU OLYMPUS CO., LTD. EVIS LUCERA ELITE COLONOVIDEOSCOPE Back to Search Results
Model Number PCF-H290ZI
Device Problems Restricted Flow rate (1248); Failure to Clean Adequately (4048)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported air/water flow issues from the air/water flow system of the evis lucera elite colonovideoscope.The device was returned for evaluation.During the device evaluation, the following reportable malfunction was identified: nozzle has foreign.There were no reports of patient or user harm associated with this event.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Manufacturer Narrative
The device was returned to olympus for evaluation and the evaluation found the nozzle had foreign objects due to insufficient cleaning.Due to clogging of nozzle, water removal ability did not meet the standard value.Additionally, the adhesive on the bending section cover had a chip and had white clouded area.The paint on the air/water and suction cylinder was peeled.The adhesive around the light guide lens had white clouded area.The connecting tube, the plastic distal end cover, and objective lens cover had scratches.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the foreign material was unable to be identified and the root cause was unable to be determined.The event can be prevented by following the instructions for use (ifu): ifu states that detection method in gif/cf/pcf-290 series operation manual chapter 3 preparation and inspection.Ifu states that preventive measure in gif/cf/pcf-290 series reprocessing manual chapter 5 reprocessing the endoscope (and related reprocessing accessories).Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA ELITE COLONOVIDEOSCOPE
Type of Device
COLONOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18246418
MDR Text Key329484602
Report Number9610595-2023-18512
Device Sequence Number1
Product Code FDF
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPCF-H290ZI
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2023
Was the Report Sent to FDA? No
Date Manufacturer Received01/31/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/02/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-