This complaint is recorded by zimmer biomet under cmp (b)(4).Following sections were updated or corrected: b4, b5, d2, d4, d9, g1, g3, g6, h1, h2, h3, h4, h6, and h11.Review of the most recent repair record determined the control bar was not in the correct position, the motor speeds were unstable, and the reciprocating arm was worn.The motor, sleeve, and reciprocating arm were replaced, and the control bar was repositioned and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|