MAUDE Adverse Event Report: COOK INC ROADRUNNER THE FIRM HYDROPHILIC WIRE GUIDE; OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ROADRUNNER THE FIRM HYDROPHILIC WIRE GUIDE; OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Catalog Number RPC-035145
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2023
Event Type  malfunction  
Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
As reported, the hydrophilic coating separated from a roadrunner the firm hydrophilic wire guide.The failure was discovered upon device receipt.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Event Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Manufacturer Narrative
Additional information: cook inc.Has been unable to submit reports due to issues with receiving new ssl certificates for esg as2 trading partners from the esg help desk.This issue was discovered over the weekend of (b)(6)2023.Cook has been in communication with the fda regarding this issue and was provided the needed security certificate on (b)(6) 2023.Per (b)(4), questions and answers about emdr - electronic medical device reporting - guidance for industry, user facilities and fda staff, section d: "¿ if a manufacturer or importer is unable to submit a report on time due to an outage affecting the esg or the cdrh emdr processing system, it may document its attempts at timely filing in block h10 for the affected reports and submit reports electronically as soon as the esg or cdrh emdr processing system is operational." an email has been sent to fda as the guidance instructs informing fda of the information set forth in the guidance regarding these delayed submissions.This report includes information known at this time.A follow-up report will be submitted should additional, relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned or that a death or serious injury occurred; nor is it admission that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Event description: as reported, the hydrophilic coating separated from a roadrunner the firm hydrophilic wire guide.The failure was discovered upon device receipt.No patient involvement.Investigation ¿ evaluation a visual inspection of the returned device was conducted.A document based investigation was also performed including a review of complaint history, device history record (dhr), the instructions for use (ifu), and quality control procedures.One roadrunner the firm hydrophilic wire guide was returned in an opened labelled package.A visual inspection noted 3cm of the inner wire was protruding from the wire guide jacket, indicating approximately 3cm of the wire guide jacket was missing.The remaining wire guide jacket measured 142cm.The tip of the jacket had a damaged appearance.A document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record found no non-conformances related to the reported failure mode.The records for incoming unpackaged coated wire guides for the relevant lots were reviewed and they all passed inspection.A review of complaint history records shows no other complaints associated with the complaint device lot.Based on the available information, cook has concluded that the device was manufactured to specification and there is no evidence suggesting nonconforming product exists either in house or in the field.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: how supplied upon removal from the package, inspect the product to ensure no damage has occurred.The cause of the complaint could not be established.Per the quality engineering risk assessment, no further action is required.The appropriate personnel have been notified, and we will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROADRUNNER THE FIRM HYDROPHILIC WIRE GUIDE
Type of Device
OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key18261101
MDR Text Key329683366
Report Number1820334-2023-01621
Device Sequence Number1
Product Code OCY
UDI-Device Identifier00827002181540
UDI-Public(01)00827002181540(17)251025(10)15036711
Combination Product (y/n)N
Reporter Country CodeMY
PMA/PMN Number
K082536
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Distributor
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 01/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/04/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberRPC-035145
Device Lot Number15036711
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received12/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-