Catalog Number ASKU
Device Problem Complete Blockage (1094)
Patient Problem Eye Burn (2523)
Event Date 11/06/2023
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported that the phaco tip was occluded and appeared ball of white material around the phaco tip, due to this phaco burn occurred and suture was used at the main incision site during the cataract [with intraocular lens (iol) implant] surgery.The surgery was completed on same day by cooling the tip with some balanced salt solution (bss).
 
Manufacturer Narrative
A sample was not received at the manufacturing site for evaluation; therefore, the condition of the product could not be verified.No lot number was identified with this complaint; therefore, a device history record review could not be conducted.A sample was not received at the manufacturing site and no lot information was provided; therefore, the root cause for the customer complaint issue cannot be determined.The cause of the reported event cannot be determined with the information obtained; therefore, specific action with regards to this complaint cannot be taken.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PHACO TIP
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18264042
MDR Text Key329674191
Report Number2523835-2023-00664
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K161794
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/04/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CENTURION VISION SYSTEM, ACCESSORY, HANDPIECE
Patient Outcome(s) Required Intervention; Other;
-
-