ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER OCUSCAN RXP OPHTHALMIC ULTRASOUND SYSTEM; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
|
Back to Search Results |
|
Catalog Number 8065741076 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A customer reported that an ophthalmic system no readings from the sensor when scanning in a-scan mode.After preventive maintenance of contacts readings appeared, but it was erroneous.Procedure details and patient impact have not been provided.Additional information has been requested.
|
|
Manufacturer Narrative
|
The system is no longer manufactured or supported by technical services.Therefore, no service was performed.The complaint investigation is being pursued at this time using available information.Based on the information available, the customer reported event cannot be confirmed.A non-conformance-based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.Based on the information obtained, the root cause of the reported event is inconclusive.Based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
|