• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. UNKNOWN 46 MAGNUM HEAD; PROTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. UNKNOWN 46 MAGNUM HEAD; PROTHESIS, HIP Back to Search Results
Device Problems Corroded (1131); Separation Failure (2547); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Cyst(s) (1800); Pain (1994); Metal Related Pathology (4530)
Event Date 05/03/2021
Event Type  Injury  
Event Description
It was reported a patient underwent a left hip revision approximately eleven years post implantation due to abdominal pain, lateral hip tenderness and, laboratory confirmed, elevated metal ion levels.During the revision, periprosthetic osteolysis in the anterosuperior region, a small posterior pseudotumor, anterosuperior bone cyst, osteolytic defect and the head was cold welded.The head and taper were removed and replaced without complication.Attempts have been made and no further information is available.
 
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2023-02879.0001825034-2023-02880.D10: unknown neck taper; unknown magnum cup.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Proposed component code: mechanical (g04) ¿ head.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Medical records/radiographs were provided and reviewed by a health care professional.A review of the available records identified findings of the reported issues.Failed left tha with periprosthetic osteolysis (anterosuperior region) and posterior pseudotumor (intracapsular ¿ posterior, quite small).Cobalt chromium levels elevated.The compliant is confirmed based on the evaluation of the provided medical records.A definitive root cause is unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN 46 MAGNUM HEAD
Type of Device
PROTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18265293
MDR Text Key329669992
Report Number0001825034-2023-02878
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/05/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexFemale
-
-