This report refers to arjo liquid flusher detergent.The report for arjo liquid flusher rinse was registered separately (under the internal record tw(b)(4)).Based on the information collected in the course of the investigation, the most likely cause of the event might be related to use error - product labeling not being followed (arjo liquid flusher detergent) intended for automatic rinsing of human waste containers was used to clean the floor.This is in line with the customer statement included in the customer report that the product was misused by trust employees.It should be underlined that the flushers are equipped with a dosing system (pump and hoses) and the dosage of arjo liquids is controlled by a microprocessor in the flusher.Following the flushers instruction for use (ifu) (e.G.For amigo flusher 60015000202_4en) : the arjo liquid flusher detergent is: "a mild alkaline detergent to be used for automated cleaning of human waste containers." "exercise caution when handling the chemical agent used in the machine.Follow the manufacturers' instructions or contact the manufacturer: if the agent comes into contact with the eyes or skin or if vapors are inhalated".In summary, the device (arjo liquid flusher detergent) was up to the manufacturer¿s specifications during the event.The product was not used for a patient hygiene at the time of the event.This complaint was decided to be reported to the competent authorities due to indication that staff members sustained serious injuries.
|