BAXTER HEALTHCARE CORPORATION MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number 5C4482 |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/07/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: the device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that there was an issue with the female connector of a minicap extended life pd transfer set.This occurred when the nurse tried to disconnect the transfer set, with the patient line (non-baxter product), from the automated peritoneal dialysis (pd) machine.The transfer set was replaced.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the sample was received for evaluation.A visual inspection with the naked eye noted a separation between the dark blue female connector and the light blue main body.Functional testing including leak, clear passage, and clamp function testing were performed with no issues noted.The reported condition was verified.The cause of the condition was related to an inadequate solvent bond between the female connector, insert chip, and the main body during the manufacturing process.A nonconformance has been opened to address this issue.A batch review was conducted and there were no deviations found related to this condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Search Alerts/Recalls
|
|
|