Catalog Number 5C4482
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/07/2023
Event Type  malfunction  
Manufacturer Narrative
H10: the device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that there was an issue with the female connector of a minicap extended life pd transfer set.This occurred when the nurse tried to disconnect the transfer set, with the patient line (non-baxter product), from the automated peritoneal dialysis (pd) machine.The transfer set was replaced.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H10: the sample was received for evaluation.A visual inspection with the naked eye noted a separation between the dark blue female connector and the light blue main body.Functional testing including leak, clear passage, and clamp function testing were performed with no issues noted.The reported condition was verified.The cause of the condition was related to an inadequate solvent bond between the female connector, insert chip, and the main body during the manufacturing process.A nonconformance has been opened to address this issue.A batch review was conducted and there were no deviations found related to this condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP TRANSFER SET
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18266683
MDR Text Key329727888
Report Number1416980-2023-06283
Device Sequence Number1
Product Code KDJ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K152675
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/05/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number5C4482
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/04/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
PATIENT LINE (NON-BAXTER).; UNSPECIFIED PD CYCLER.
-
-