Model Number 801188
Device Problem Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/10/2023
Event Type  malfunction  
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the central control monitor (ccm) displayed a 'service gas system alarm' message.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The field service representative (fsr) replaced the oxygen (o2) sensor and performed release testing.The unit operated to the manufacturer's specifications.
 
Manufacturer Narrative
The field service representative (fsr) replaced the oxygen (o2) sensor and performed release testing.The unit operated to the manufacturer's specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key18271099
MDR Text Key329807110
Report Number1828100-2023-00364
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup,Followup
Report Date 04/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/05/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801188
Device Catalogue Number801188
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/16/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-