Catalog Number 320109 |
Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
B.3.Date of event: unknown.The date received by manufacturer has been used for this field.H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that the bd ultra-fine¿ short pen needles cap does not attach.The following was received from the initial reporter: consumer reported finding the needle covers falling off the needle.Caller is storing the needle on the pen in his travel case date of event unknown date of event.
|
|
Manufacturer Narrative
|
H.6.Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time h3 other text : see h.10.
|
|
Event Description
|
It was reported that the bd ultra-fine¿ short pen needles cap does not attach.The following was received from the initial reporter: consumer reported finding the needle covers falling off the needle.Caller is storing the needle on the pen in his travel case.Date of event unknown date of event.
|
|