It was reported that the cardiohelp would alarm ¿venous bubble sensor is defective¿ and the customer switched the sensor and used one from another device.The failure occurred during treatment.A getinge field service technician (fst) was sent for investigation and repair.The venous bubble sensor was replaced.The fst performed safety, calibration, and functionality checks to factory specifications.All function tests are passed.The log files of the reported cardiohelp device were reviewed and the error message "venous bubble sensor is defective" could be confirmed on the date of event.No exact root cause could be determined.However, another venous bubble sensor with a similar failure was already investigated by the supplier sonotec.Following possible root causes were determined: - damaged wiring inside the cable due to mechanical tension - damage due to overvoltage or esd (electrostatic discharge) according to the instructions for use (ifu), chapters 2.2.5 "monitoring and sensors" and 5.4.4 "bubble monitoring: function test" the venous bubble sensor and the arterial flow/bubble sensor have to be tested before each use.The cardiohelp has a flow/bubble sensor for bubble detection.The venous bubble sensor is optional and for additional bubble detection.The device was manufactured on 2022-10-18.The device history record (dhr) of the cardiohelp (material: 701072780, serial: (b)(6) was reviewed on 2023-10-09.There is no indication that manufacturing issues occurred, thus production related influences are unlikely to have contributed to the reported failure.Based on the results the reported failure "venous bubble sensor is defective" could be confirmed.The customer will be informed about the results by the getinge sales and service unit.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.
|