Model Number CARDIOHELP
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/06/2023
Event Type  malfunction  
Manufacturer Narrative
It was reported that the cardiohelp would alarm ¿venous bubble sensor is defective¿ and the customer switched the sensor and used one from another device.The failure occurred during treatment.A getinge field service technician (fst) was sent for investigation and repair.The venous bubble sensor was replaced.The fst performed safety, calibration, and functionality checks to factory specifications.All function tests are passed.The log files of the reported cardiohelp device were reviewed and the error message "venous bubble sensor is defective" could be confirmed on the date of event.No exact root cause could be determined.However, another venous bubble sensor with a similar failure was already investigated by the supplier sonotec.Following possible root causes were determined: - damaged wiring inside the cable due to mechanical tension - damage due to overvoltage or esd (electrostatic discharge) according to the instructions for use (ifu), chapters 2.2.5 "monitoring and sensors" and 5.4.4 "bubble monitoring: function test" the venous bubble sensor and the arterial flow/bubble sensor have to be tested before each use.The cardiohelp has a flow/bubble sensor for bubble detection.The venous bubble sensor is optional and for additional bubble detection.The device was manufactured on 2022-10-18.The device history record (dhr) of the cardiohelp (material: 701072780, serial: (b)(6) was reviewed on 2023-10-09.There is no indication that manufacturing issues occurred, thus production related influences are unlikely to have contributed to the reported failure.Based on the results the reported failure "venous bubble sensor is defective" could be confirmed.The customer will be informed about the results by the getinge sales and service unit.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
It was reported that the device alarmed ¿venous bubble sensor defective¿ but the pump did not stop.No harm to any person has been reported.Complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOHELP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key18272818
MDR Text Key329777007
Report Number8010762-2023-00600
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCARDIOHELP
Device Catalogue Number701072780
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/30/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/18/2022
Is the Device Single Use? No
Type of Device Usage Initial
Patient Sequence Number1
-
-