Model Number 60-05-40 |
Device Problem
Difficult to Open or Close (2921)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/09/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A.1.-a.5.Patient information was not provided.H10: livanova deutschland manufactures the s5 erc tubing clamp.The incident occurred in usa.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report that, during a prcedure, an erc malfunction was reported.There is no report of any patient injury.
|
|
Manufacturer Narrative
|
H10: a livanova field service representative was dispatched to the facility to investigate the device: the erc tubing clamp was working properly.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
See initial report.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H10: according to sap installed base analysis for involved customer, possible erc affected could be model 60-05-40, serial number (b)(6).Model and serial number have been added to the dedicated d.4 section.Based on technical intervention carried out, a hardware malfunction with the clamp can be ruled out.Therefore, it cannot be excluded that the reported event could have been caused by an incorrect set system by the user.
|
|