An olympus personnel reported on behalf of the customer, defects of the bending section and insertion tube of the airway mobilescope.The device was returned for evaluation.During the device evaluation, the following reportable malfunction was identified: connecting tube has coating peeling.(1mm2 or more).There were no reports of patient or user harm associated with this event.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
|
The device was returned to olympus for evaluation and the evaluation found the connecting tube had coating peeling.(1mm2 or more) and a dent damage.Additionally, due to pinhole damage on the connecting tube and the bending section cover, water tightness was lost.Due to damage on image guide bundle, the image has a stain.The camera section had a scratch and a crack.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event is unable to be determined.However, the cause of the event is likely due to stress of repeated use, external factors, or handling.The event can be detected by following the instructions for use (ifu) sections which state: there is a warning as below in the instruction manual ¿chapter 3 preparation and inspection 3.6 inspection of the endoscope".Inspection of the endoscope 4.Inspect the boot and the insertion section near the boot for bends, twists, or other irregularities.5.Inspect the external surface of the entire insertion section including the bending section and the distal end for dents, bulges, swelling, scratches, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities.6.Holding the control section with one hand, carefully run your other hand back and forth over the entire length of the insertion section confirm that no objects or metallic wire protrudes from the insertion section.Also, confirm that the insertion tube is not abnormally rigid.Olympus will continue to monitor field performance for this device.
|