AIZU OLYMPUS CO., LTD. AIRWAY MOBILESCOPE Back to Search Results
Model Number MAF-TM
Device Problems Peeled/Delaminated (1454); Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
An olympus personnel reported on behalf of the customer, defects of the bending section and insertion tube of the airway mobilescope.The device was returned for evaluation.During the device evaluation, the following reportable malfunction was identified: connecting tube has coating peeling.(1mm2 or more).There were no reports of patient or user harm associated with this event.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Manufacturer Narrative
The device was returned to olympus for evaluation and the evaluation found the connecting tube had coating peeling.(1mm2 or more) and a dent damage.Additionally, due to pinhole damage on the connecting tube and the bending section cover, water tightness was lost.Due to damage on image guide bundle, the image has a stain.The camera section had a scratch and a crack.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event is unable to be determined.However, the cause of the event is likely due to stress of repeated use, external factors, or handling.The event can be detected by following the instructions for use (ifu) sections which state: there is a warning as below in the instruction manual ¿chapter 3 preparation and inspection 3.6 inspection of the endoscope".Inspection of the endoscope 4.Inspect the boot and the insertion section near the boot for bends, twists, or other irregularities.5.Inspect the external surface of the entire insertion section including the bending section and the distal end for dents, bulges, swelling, scratches, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities.6.Holding the control section with one hand, carefully run your other hand back and forth over the entire length of the insertion section confirm that no objects or metallic wire protrudes from the insertion section.Also, confirm that the insertion tube is not abnormally rigid.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRWAY MOBILESCOPE
Type of Device
AIRWAY MOBILESCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18276981
MDR Text Key329929277
Report Number9610595-2023-18925
Device Sequence Number1
Product Code EOQ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K082720
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMAF-TM
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2023
Was the Report Sent to FDA? No
Date Manufacturer Received12/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/02/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse